A Feasibility Study to Predict 3D Dose Delivery Accuracy for IMRT Using DenseNet with Log Files
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY(2024)
Fudan Univ
Abstract
OBJECTIVE: This study aims to explore the feasibility of DenseNet in the establishment of a three-dimensional (3D) gamma prediction model of IMRT based on the actual parameters recorded in the log files during delivery. METHODS: A total of 55 IMRT plans (including 367 fields) were randomly selected. The gamma analysis was performed using gamma criteria of 3% /3 mm (Dose Difference/Distance to Agreement), 3% /2 mm, 2% /3 mm, and 2% /2 mm with a 10% dose threshold. In addition, the log files that recorded the gantry angle, monitor units (MU), multi-leaf collimator (MLC), and jaws position during delivery were collected. These log files were then converted to MU-weighted fluence maps as the input of DenseNet, gamma passing rates (GPRs) under four different gamma criteria as the output, and mean square errors (MSEs) as the loss function of this model. RESULTS: Under different gamma criteria, the accuracy of a 3D GPR prediction model decreased with the implementation of stricter gamma criteria. In the test set, the mean absolute error (MAE) of the prediction model under the gamma criteria of 3% /3 mm, 2% /3 mm, 3% /2 mm, and 2% /2 mm was 1.41, 1.44, 3.29, and 3.54, respectively; the root mean square error (RMSE) was 1.91, 1.85, 4.27, and 4.40, respectively; the Sr was 0.487, 0.554, 0.573, and 0.506, respectively. There was a correlation between predicted and measured GPRs (P < 0.01). Additionally, there was no significant difference in the accuracy between the validation set and the test set. The accuracy in the high GPR group was high, and the MAE in the high GPR group was smaller than that in the low GPR group under four different gamma criteria. CONCLUSIONS: In this study, a 3D GPR prediction model of patient-specific QA using DenseNet was established based on log files. As an auxiliary tool for 3D dose verification in IMRT, this model is expected to improve the accuracy and efficiency of dose validation.
MoreTranslated text
Key words
Three-dimensional gamma prediction,log files,patient-specific quality assurance
求助PDF
上传PDF
View via Publisher
AI Read Science
AI Summary
AI Summary is the key point extracted automatically understanding the full text of the paper, including the background, methods, results, conclusions, icons and other key content, so that you can get the outline of the paper at a glance.
Example
Background
Key content
Introduction
Methods
Results
Related work
Fund
Key content
- Pretraining has recently greatly promoted the development of natural language processing (NLP)
- We show that M6 outperforms the baselines in multimodal downstream tasks, and the large M6 with 10 parameters can reach a better performance
- We propose a method called M6 that is able to process information of multiple modalities and perform both single-modal and cross-modal understanding and generation
- The model is scaled to large model with 10 billion parameters with sophisticated deployment, and the 10 -parameter M6-large is the largest pretrained model in Chinese
- Experimental results show that our proposed M6 outperforms the baseline in a number of downstream tasks concerning both single modality and multiple modalities We will continue the pretraining of extremely large models by increasing data to explore the limit of its performance
Upload PDF to Generate Summary
Must-Reading Tree
Example

Generate MRT to find the research sequence of this paper
Data Disclaimer
The page data are from open Internet sources, cooperative publishers and automatic analysis results through AI technology. We do not make any commitments and guarantees for the validity, accuracy, correctness, reliability, completeness and timeliness of the page data. If you have any questions, please contact us by email: report@aminer.cn
Chat Paper
Summary is being generated by the instructions you defined