The Impact of Mcs Models and Efac Values on the Dose Simulation for A Proton Pencil Beam

RADIATION PHYSICS AND CHEMISTRY(2017)

Chang Gung Univ

Cited 3|Views33
Abstract
The Multiple Coulomb Scattering (MCS) model plays an important role in accurate MC simulation, especially for small field applications. The Rossi model is used in MCNPX 2.7.0, and the Lewis model in Geant4.9.6.p02. These two models may generate very different angular and spatial distributions in small field proton dosimetry. Beside angular and spatial distributions, step size is also an important issue that causes path length effects. The Energy Fraction (EFAC) value can be used in MCNPX 2.7.0 to control step sizes of MCS. In this study, we use MCNPX 2.7.0, Geant4.9.6.p02, and one pencil beam algorithm to evaluate the effect of dose deposition because of different MCS models and different EFAC values in proton disequilibrium situation. Different MCS models agree well with each other under a proton equilibrium situation. Under proton disequilibrium situations, the MCNPX and Geant4 results, however, show a significant deviation (up to 43%). In addition, the path length effects are more significant when EFAC is equal to 0.917 and 0.94 in small field proton dosimetry, and using a 0.97 EFAC value is the best for both accuracy and efficiency
More
Translated text
Key words
Small field,Monte Carlo,Proton therapy,Multiple Coulomb scattering,Path length correction
PDF
Bibtex
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Data Disclaimer
The page data are from open Internet sources, cooperative publishers and automatic analysis results through AI technology. We do not make any commitments and guarantees for the validity, accuracy, correctness, reliability, completeness and timeliness of the page data. If you have any questions, please contact us by email: report@aminer.cn
Chat Paper
Summary is being generated by the instructions you defined